The High-Frequency High-Voltage Insulation Tester / High-Frequency Dielectric Strength Tester (HFIT8.0) is designed in accordance with IEC60601-2-2:2017 (GB 9706.202-2021) and is used for testing the high-frequency leakage current (201.8.8.3.102a) and high-frequency dielectric strength (201.8.8.3.103) of high-frequency electrosurgical accessories. The HFIT8.0 can output multiple waveforms, with a maximum test voltage of up to 7200Vp and an adjustable crest factor of 1.4–7.0, covering the testing requirements for the vast majority of high-frequency electrosurgical accessories on the market.
Product Features
1. Multiple output waveforms: Sine, Modified Sine, Burst, Pulse;
2. Maximum output voltage up to 7200Vp, with an adjustable crest factor from 1.4 to 7.0;
3. Automatic mode: users only need to set the test voltage, and the machine automatically selects the output waveform and adjusts the crest factor;
4. Built-in timer function, allowing the test duration to be set;
5. Built-in monitoring values for voltage, current, crest factor, frequency, etc.;
6. Provides an oscilloscope monitoring port, allowing direct connection to an oscilloscope to monitor voltage and current without the need for additional voltage and current probes;
Technical Parameters
1. Output modes: Sine, Modified Sine, Burst, Pulse, Auto;
2. Output voltage: 200–7200Vp, accuracy ±3%±4V;
3. Crest factor: 1.4–7.0;
4. Leakage current: 0–350mArms, accuracy ±3%±2mArms;
5. Oscilloscope port: Voltage 1000:1, Current 1:1