I. Overview

The insulation resistance meter (megohmmeter) is a nationally mandatory verification measuring instrument, with its verification regulation being JJG622-89. The main standard devices of the verification device are the high-voltage high-resistance box and the high-voltage voltmeter. The GZX92A insulation resistance meter verification device integrates the high-voltage high-resistance box (high-resistance output part) and the high-voltage digital voltmeter (voltage input part) into one unit.

The measurement upper limit of the resistance output part in the verification device is 200 GΩ, with a nominal working voltage of 5 kV. The device is designed with independent leakage terminals and grounding terminals, showing no significant instability, short-circuit, or open-circuit phenomena during measurement. It meets the requirements of the JJG622-97 insulation resistance meter verification regulation and the JJG166-93 DC resistor verification regulation. It can be used for the verification of various models of pointer-type insulation resistance meters.

The resistance output part of the verification device uses a patented rolling high-voltage switch, which can still meet the verification regulation requirements after 100,000 uses. It has a long technical and mechanical lifespan. The nominal measurement voltage of the voltage measurement part in the verification device is 5 kV, with an input impedance of ≥10 GΩ. This device can measure the open-circuit voltage, median voltage, and peak voltage of the megohmmeter.

II. Main Technical Parameters

1. Resistance Output (High-Voltage High-Resistance Box) Part

1.1 Accuracy Class and Working Voltage (Current) of the Resistance Output Part

Resistance Value

100 GΩ

×10 GΩ

×1 GΩ

×100 MΩ

×10 MΩ

×1 MΩ

Accuracy Class

5

5

2

1

0.5

0.2

Nominal Voltage

5000 V

5000 V

5000 V

5000 V

2500 V

1000 V

Resistance Value

×100 kΩ

×10 kΩ

×1 kΩ

×100 Ω

Accuracy Class

0.2

0.2

0.2

0.2

Nominal Current

1 mA

8 mA

20 mA

50 mA

1.2 Adjustment Range: 100 Ω–200 GΩ, with a fine adjustment of 100 Ω

1.3 Operating Environmental Conditions

1.3.1 Reference Temperature Range: 20–25°C

1.3.2 Nominal Operating Temperature Range: 18–28°C

1.3.3 Reference Humidity Range: 40–60%

1.3.4 Nominal Operating Humidity Range: 25–75%

1.4 Resistance Variation Limit: Under reference conditions, the variation caused by a change in a single influencing factor

Influencing Factor

Nominal Operating Range

Permissible Variation

Ambient Temperature

18–28°C

a/2 %

Relative Humidity

25%–75%

≤10 GΩ a/2 %; >10 GΩ a%

Operating Voltage

(1–1/5) Nominal Voltage

≤10 GΩ a/2 %; >10 GΩ a%

★ a is the accuracy class of the resistance dial in the resistance output part of the verification device (a values differ for each resistance dial).

1.5 Insulation Resistance: The insulation resistance between the circuit of the resistance output part of the verification device and any other external metal without electrical connection. The resistance value measured at the nominal voltage is not less than 5 TΩ.

1.6 Insulation Strength: Between the circuit of the resistance output part of the verification device and the reference grounding point for testing, it should withstand an actual sinusoidal AC voltage of 11 kV at a frequency of 45–65 Hz for 1 minute without breakdown or flashover.

1.7 The residual resistance at the output terminals of the resistance output part of the verification device should be <0.1 Ω, with a variation of <0.01 Ω.

1.8 Dimensions: 442 mm × 270 mm × 145 mm

1.9 Weight: <5 kg

2. Voltage Measurement (High-Voltage DC Digital Voltmeter) Part

2.1 Measurement Range: 0–5500 V

2.2 Accuracy: ±(1% of reading ±1 digit)

2.3 Power Supply: DC 9 V

2.4 Input Resistance: ≥10 GΩ

2.5 Display: 4.5-digit, LCD display

2.6 The peak voltage measurement circuit meets the requirements of Figure 3 in the JJG622-97 insulation resistance meter verification regulation.


Industry Category
Product Category
Brand: 深圳世紀(jì)經(jīng)典
Spec: GZX92A型絕緣電阻表檢定裝置
Stock:
Origin: China / Guangdong / Shenshi
About Toocle.com - Partner Programme - Old Version
Copyright ? Toocle.com. All Rights Reserved.
(浙)-經(jīng)營性-2023-0192
ChatGlobal Chat Now 若尔盖县| 肥乡县| 昌江| 石嘴山市| 集贤县| 白河县| 景德镇市| 博客| 五常市| 剑川县| 海宁市| 佛冈县| 阿鲁科尔沁旗| 新龙县| 乌鲁木齐市| 杭州市| 嵊泗县| 隆尧县| 二连浩特市| 绵竹市| 呼和浩特市| 潜江市| 尤溪县| 博野县| 高台县| 湖南省| 东阿县| 广南县| 阳泉市| 潜江市| 吴忠市| 松溪县| 静宁县| 望江县| 安岳县| 西畴县| 南乐县| 锡林郭勒盟| 汉沽区| 台东市| 嘉义县|